Assay Type
Solid Phase Sandwich ELISA
Format
96-well strip plate
Assay Length
4.5 hours
Sample Type & Volume Required Per Well
Cell Culture Supernates (50 uL), Serum (30 uL), EDTA Plasma (30 uL), Heparin Plasma (30 uL), Urine (30 uL)
Sensitivity
5.1 pg/mL
Assay Range
31.2 - 2,000 pg/mL (Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Urine)
Specificity
Natural and recombinant human CD23. This assay recognizes the 25 kDa fragment (aa 150-321) and the full length extracellular domain (aa 48-321) of human CD23/Fc epsilon RII.
Cross-reactivity
< 0.5% cross-reactivity observed with available related molecules.Cross-species reactivity not tested.
Interference
No significant interference observed with available related molecules.

Product Summary

The Quantikine Human CD23 immunoassay is a 4.5 hour solid phase ELISA designed to measure human soluble CD23 in cell culture supernates, serum, plasma, and urine. It contains NS0-expressed recombinant human CD23 (extracellular domain) and has been shown to accurately quantitate the recombinant factor. Results obtained using natural human CD23 showed linear curves that were parallel to the standard curves obtained using the Quantikine kit standards. These results indicate that this kit can be used to determine relative mass values for naturally occurring soluble CD23.

Precision

Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision
Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision

Cell Culture Supernates, Serum, EDTA Plasma, Heparin Plasma, Urine

Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 40 40 40
Mean (pg/mL) 158 283 576 151 274 533
Standard Deviation 13.5 17.8 40.8 13.7 22.4 40.2
CV% 8.5 6.3 7.1 9.1 8.2 7.5

Recovery

The recovery of CD23 spiked to levels throughout the range of the assay in various matrices was evaluated.

Sample Type Average % Recovery Range %
Cell Culture Media (n=4) 105 94-114
EDTA Plasma (n=4) 99 89-105
Heparin Plasma (n=4) 98 90-112
Serum (n=4) 94 85-108
Urine (n=4) 100 85-113

Linearity

To assess the linearity of the assay, samples containing high concentrations of CD23 were serially diluted with Calibrator Diluent to produce samples with values within the dynamic range of the assay.
Human CD23/Fc epsilon RII ELISA Linearity

Scientific Data

Human CD23/Fc epsilon RII ELISA Standard Curve

Product Datasheets

You must select a language.

x

Preparation and Storage

Shipping
The product is shipped at ambient temperature. Upon receipt, store it immediately at the temperature recommended below.
Storage
Store the unopened product at 2 - 8 °C. Do not use past expiration date.

Background: CD23/Fc epsilon RII

CD23, also known as Fc epsilon RII, is a type 2 transmembrane C-type lectin that binds IgE, CD21, CD11b and CD11c. It is expressed on a variety of hematopoietic cells and functions in allergic immune responses.

Long Name:
Fc epsilon Receptor II
Entrez Gene IDs:
2208 (Human); 14128 (Mouse); 171075 (Rat)
Alternate Names:
BLAST-2; CD23; CD23A; CD23CD23 antigen; CLEC4J; CLEC4JC-type lectin domain family 4 member J; C-type lectin domain family 4, member J; Fc epsilon RII; Fc fragment of IgE, low affinity II, receptor for (CD23); FCE2Fc fragment of IgE, low affinity II, receptor for (CD23A); fc-epsilon-RII; FCER2; Fcer2a; FceRII; IGEBF; Immunoglobulin E-binding factor; low affinity immunoglobulin epsilon Fc receptor; Ly-42; Lymphocyte IgE receptor
⚠ WARNING: This product can expose you to chemicals including N,N-Dimethylforamide, which is known to the State of California to cause cancer. For more information, go to www.P65Warnings.ca.gov.

Assay Procedure

Refer to the product for complete assay procedure.

Bring all reagents and samples to room temperature before use. It is recommended that all samples, standards, and controls be assayed in duplicate.
  1.   Prepare all reagents, standard dilutions, and samples as directed in the product insert.
  2.   Remove excess microplate strips from the plate frame, return them to the foil pouch containing the desiccant pack, and reseal.

  3. 50 µL Assay Diluent
  4.   Add 50 µL of Assay Diluent to each well.

  5. 50 µL Standard, Control, or Sample
  6.   Add 50 µL of Standard, control, or sample to each well. Cover with a plate sealer, and incubate at room temperature for 3 hours.
  7.   Aspirate each well and wash, repeating the process 3 times for a total of 4 washes.

  8. 100 µL Conjugate
  9.   Add 100 µL of Conjugate to each well. Cover with a new plate sealer, and incubate at room temperature for 1 hour.
  10.   Aspirate and wash 4 times.

  11. 100 µL Substrate Solution
  12.   Add 100 µL Substrate Solution to each well. Incubate at room temperature for 30 minutes. PROTECT FROM LIGHT.

  13. 100 µL Stop Solution
  14.   Add 100 µL of Stop Solution to each well. Read at 450 nm within 30 minutes. Set wavelength correction to 540 nm or 570 nm.

Citation for Human CD23/Fc epsilon RII Quantikine ELISA Kit

R&D Systems personnel manually curate a database that contains references using R&D Systems products. The data collected includes not only links to publications in PubMed, but also provides information about sample types, species, and experimental conditions.

1 Citation: Showing 1 - 1

  1. Mediating effect of soluble B-cell activation immune markers on the association between anthropometric and lifestyle factors and lymphoma development
    Authors: F Saberi Hos, PM Kolijn, D Casabonne, A Nieters, M Solans, S Naudin, P Ferrari, JD Mckay, E Weiderpass, V Perduca, C Besson, FR Mancini, G Masala, V Krogh, F Ricceri, JM Huerta, D Petrova, N Sala, A Trichopoul, A Karakatsan, C La Vecchia, R Kaaks, F Canzian, D Aune, H Boeing, MB Schulze, A Perez-Corn, AW Langerak, VHJ van der Ve, R Vermeulen
    Sci Rep, 2020-08-14;10(1):13814.
    Species: Human
    Sample Types: Serum

FAQs

No product specific FAQs exist for this product, however you may

View all ELISA FAQs
Loading...

Reviews for Human CD23/Fc epsilon RII Quantikine ELISA Kit

There are currently no reviews for this product. Be the first to review Human CD23/Fc epsilon RII Quantikine ELISA Kit and earn rewards!

Have you used Human CD23/Fc epsilon RII Quantikine ELISA Kit?

Submit a review and receive an Amazon gift card.

$25/€18/£15/$25CAN/¥75 Yuan/¥2500 Yen for a review with an image

$10/€7/£6/$10 CAD/¥70 Yuan/¥1110 Yen for a review without an image

Submit a Review