Human EMMPRIN/CD147 Magnetic Luminex® Performance Assay

Build your own Luminex Assay with our Luminex Assay Customization Tool.

U.S. orders usually ship in 2-3 business days

Build Now
Quantist software graph of signal intensity versus concentration for EMMPRIN/CD147 in the Human MMP Luminex Panel
1 Image
Product Details
Citations (2)
FAQs
Supplemental Products
Reviews

Human EMMPRIN/CD147 Magnetic Luminex® Performance Assay Summary

Panel Luminex Performance Human MMP Panel
Assay Type Magnetic bead-based multiplex assay for the Luminex® platform
Format 1 x 96-well microplate and Magnetic antibody-coated beads
Species Human
Analytes Detected Please see analyte list in assay customization tool below.
Performance Validation Our Luminex High Performance Assays undergo our most extensive validation testing. All Luminex High Performance Assay Panels are validated for use with serum and plasma samples. Some High Performance Assays are additionally validated for cell culture supernatants, milk, saliva, or urine, as indicated on the individual product data sheets. Luminex High Performance Assays are tested for sensitivity (three-quarters the low standard), intra-assay precision, inter-assay precision, and to ensure assay linearity for validated sample types. Recovery values for individual samples in validated sample types are also tested. Validation data for each analyte can be found in the product datasheet.

Make your selections below to configure your assay

Filter Analytes
Select All Select None

Assays

Below are kits that are available based on your selections. These selections may be further customized before adding to the cart or emailing the configuration to your purchasing department or local distributor to place your order. Selected analytes may be included in more than one High Performance Assay. All assay options are displayed.

If you run into bead region incompatibilities, send your Luminex code to our team and find out more.

How would you like these analytes mixed?

Premixed
User Mixed
Please select at least one mixing option to see the kits.

Would you like to split based on dilution?

Do not split
Split based on serum
Split based on plasma
Split based on cell culture supernates

Not finding your analyte of interest? View our complete Luminex offering in the full version of our Luminex Assay Customization Tool. We also offer Custom Luminex Services.

Product Features

Magnetic Luminex High Performance Assays are flexible bead-based multiplex assays. They allow up to 50 user-defined target analytes to be simultaneously profiled using cell culture supernates, serum, or plasma samples. Some panels are also validated with human milk, saliva, and/or urine samples.

  • Simultaneously profile Human EMMPRIN/CD147 with up to analytes of your choice in one sample
  • Magnetic format allows for easier wash steps
  • Undergoes similar validation testing as Quantikine® ELISA Kits
  • Precise and Specific
  • Requires a small sample volume (<50 µL)
  • Run your assay in just 3-5 hours
  • Mass-calibrated standards for consistent results with every new lot of material
  • Enough reagents for one 96-well plate
Base Kit Components
  • Standard Cocktail(s)
  • Bead Diluent
  • Biotin Antibody Diluent
  • Calibrator Diluent
  • Wash Buffer
  • Streptavidin-PE
  • One flat-bottom 96-well Microplate
  • Foil Plate Sealers (4)
  • Mixing Bottle
  • Standard Value Card**

**A standard curve must be generated each time an assay is run, utilizing values from the Standard Value Card included in the Base Kit.

Bead Set Contents
  • User-mix antibody-coated Magnetic beads in individual vials
  • User-mixed cocktail of biotinylated detection antibody

Precision

Intra-assay Precision (precision within an assay) - Three samples of known concentration were tested twenty times on one plate to assess intra-assay precision.
Inter-assay Precision (precision between assays) - Three samples of known concentration were tested in seventy-two separate assays to assess precision between assays.

  Intra-Assay Precision Inter-Assay Precision
Sample 1 2 3 1 2 3
n 20 20 20 72 72 72
Mean (pg/mL) 265 1785 7075 256 1782 7284
Standard deviation 19.6 109 424 29.2 152 696
CV (%) 7.4 6.1 6.0 11.4 8.5 9.5
 

Recovery

Samples containing and/or spiked with human EMMPRIN were evaluated for recovery.

Sample Type Average % Recovery Range (%)
Cell culture supernate 96 72-123
 

Linearity

Samples containing and/or spiked with human EMMPRIN were serially diluted to evaluate assay linearity.

  Cell culture supernate Serum Heparin plasma Platelet-poor heparin plasma Urine
1:2 Average % of Expected 114 86 99 93 91
Range (%) 103-122 79-91 98-99 78-105 89-93
1:4 Average % of Expected 114 82 95 91 82
Range (%) 98-123 80-86 90-101 85-97 76-87
1:8 Average % of Expected 111 78 99 87 82
Range (%) 97-121 75-81 96-107 77-91 77-90
Which Luminex Assay is right for me?

Assays for the Luminex platform are offered as High Performance Assays or Assays. The Luminex High Performance Assays are fully validated panels with a focused selection of analytes. They are available in We Mix, You Mix, and Predetermined formats. The Luminex Assays allows for the maximum number of analytes in a multiplex and is supplied as a premixed kit. View a table comparing the features and benefits of these bead-based multiplex assays.

More Information About Magnetic Bead-based Multiplex Kits For The Luminex® Platform

Magnetic Luminex High Performance Assays are designed for use with the Luminex MAGPIX CCD Imager. Alternatively, kits can be used with the Luminex 100/200™ or FLEXMAP 3D, dual laser, flow-based sorting and detection platforms.

Analyte-specific antibodies are pre-coated onto color-coded microparticles. Microparticles, standards, and samples are pipetted into wells and the immobilized antibodies bind the analytes of interest. After washing away any unbound substances, a biotinylated antibody cocktail specific to the analytes of interest is added to each well. Following a wash to remove any unbound biotinylated antibody, Streptavidin-Phycoerythrin conjugate (Streptavidin-PE), which binds to the biotinylated detection antibodies, is added to each well. A final wash removes unbound Streptavidin-PE and the microparticles are resuspended in buffer and read using the Luminex or Bio-Plex analyzer. One laser is bead-specific and determines which analyte is being detected. A magnet in the analyzer captures and holds the superparaMagnetic microparticles in a monolayer. Two spectrally distinct Light Emitting Diodes (LEDs) illuminate the beads. One LED identifies the analyte that is being detected and the second LED determines the magnitude of the PE-derived signal, which is in direct proportion to the amount of analyte bound. Each well is imaged with a CCD camera. Kits can also be used with Luminex 100/200 or a Bio-Rad Bio-Plex dual laser, flow-based systems.

Scientific Data

Quantist software graph of signal intensity versus concentration for EMMPRIN/CD147 in the Human MMP Luminex Panel

Background: EMMPRIN/CD147

The name EMMPRIN reflects this proteins extracellular metalloproteinase inducer activity. It is a cell surface protein with two extracellular Ig domains, a transmembrane domain, and a cytoplasmic domain. Its expression is often elevated in human tumor cells, which in turn promotes MMP-1, MMP-2 and MMP-3 production in melanoma cells and MMP-2 and MMP-9 production in breast cancer cells. It interacts with several types of molecules such as integrins, caveolin-1 and MCT1. TRA-1-85 is an epitope within EMMPRIN.

Long Name
Extracellular Matrix Metalloproteinase Inducer
Entrez Gene IDs
682 (Human); 12215 (Mouse); 102121721 (Cynomolgus Monkey)
Alternate Names
5F7; basigin (Ok blood group); Basigin; BSG; CD147 antigen; CD147; Collagenase stimulatory factor; EMMPRIN; EMMPRINTCSF; Extracellular matrix metalloproteinase inducer; Leukocyte activation antigen M6; M6; OK blood group antigen; OK; Tumor cell-derived collagenase stimulatory factor

Product Datasheets

You must select a language.

x
Product Specific Notices
Luminex® is a registered trademark of Luminex Corporation.

Citations for Human EMMPRIN/CD147 Magnetic Luminex® Performance Assay

R&D Systems personnel manually curate a database that contains references using R&D Systems products. The data collected includes not only links to publications in PubMed, but also provides information about sample types, species, and experimental conditions.

2 Citations: Showing 1 - 2
Filter your results:

Filter by:

  1. Melanoma Cell Resistance to Vemurafenib Modifies Inter-Cellular Communication Signals
    Authors: C Tabolacci, M Cordella, S Mariotti, S Rossi, C Senatore, C Lintas, L Levati, D D'Arcangel, A Facchiano, S D'Atri, R Nisini, F Facchiano
    Biomedicines, 2021-01-15;9(1):.
    Species: Human
    Sample Types: Cell Culture Supernates
  2. The impact of human adipose tissue-derived stem cells on breast cancer cells: implications for cell-assisted lipotransfers in breast reconstruction
    Authors: E Koellenspe, LC Bonnert, I Zoernig, F Marmé, S Sandmann, G Germann, F Gramley, U Leimer
    Stem Cell Res Ther, 2017-05-25;8(1):121.
    Species: Human
    Sample Types: Cell Culture Supernates

FAQs

No product specific FAQs exist for this product, however you may

View all Luminex Assay FAQs
Loading...

Reviews for Human EMMPRIN/CD147 Magnetic Luminex® Performance Assay

There are currently no reviews for this product. Be the first to review Human EMMPRIN/CD147 Magnetic Luminex® Performance Assay and earn rewards!

Have you used Human EMMPRIN/CD147 Magnetic Luminex® Performance Assay?

Submit a review and receive an Amazon gift card.

$25/€18/£15/$25CAN/¥75 Yuan/¥2500 Yen for a review with an image

$10/€7/£6/$10 CAD/¥70 Yuan/¥1110 Yen for a review without an image

Submit a Review