Troubleshooting Guide: Human Proteome Profiler Array Kits

For further assistance, please contact our technical service department.

Observation Problem Corrective Action
No signals on positive control spots Concentration of detection antibody and/or SA-HRP too low Use concentration/dilution specified
Chemiluminescent reagents failed Repeat with fresh chemiluminescent reagents according to manufacturers protocol
Inadequate exposure time to film Increase exposure time
No or low signals on target spots Sample dilution too high Use more sample
Analyte abundance in sample is low Use more sample
Use different cell line
Verify that conditions used to stimulate cells were optimal
Sample deteriorated during preparation Supplement buffers with protease and phosphatase inhibitors where indicated
Sample deteriorated during storage All samples should be stored at ≤ -70 °C.
Avoid freeze-thaw cycles
Inadequate exposure time to film Increase exposure time
Signals on negative control spots Concentration of detection antibody and/or SA-HRP too high Use concentration/dilution specified
Sample concentration too high Use less sample
Uneven or high background on blank areas of array Insufficient washing Perform number of washes with volume indicated as specified in the product insert protocol
Perform washes in large container and not the 4-Well Multi-dish
Concentration of detection antibody and/or SA-HRP too high Use concentration/dilution specified
Array was allowed to dry out partially during the procedure Always keep arrays submerged. Minimize time the array is exposed to air.
Creased or scratched arrays Use flat-tipped forceps to handle arrays on array number.
Too many spots have signals Insufficient washing Perform washes as stated in product insert. Perform washes in a large container, not in the 4-Well Multi-dish.
Concentration of detection antibody and/or SA-HRP too high Use concentration/dilution specified
Sample concentration too high Use less sample